pmcSensors (Basel)Sensors (Basel)1660sensors101204366sensorsSensors (Basel, Switzerland)1424-8220Multidisciplinary Digital Publishing Institute (MDPI)PMC9268973PMC9268973.1926897392689733580816310.3390/s22134668sensors-22-046681ArticlePossibilities of Real Time Monitoring of Micropollutants in Wastewater Using Laser-Induced Raman & Fluorescence Spectroscopy (LIRFS) and Artificial Intelligence (AI)https://orcid.org/0000-0003-4387-4712PostClaudia1*HeydenNiklas1*ReinartzAndré1FoerdererAaron2BruelisauerSimon3https://orcid.org/0000-0003-1157-976XLinnemannVolker4https://orcid.org/0000-0002-0403-9979HugWilliam5AmannFlorian1MichaelidesMichalisAcademic EditorDepartment of Engineering Geology and Hydrogeology, RWTH Aachen University, Lochnerstr. 4-20, 52064 Aachen, Germany; andre.reinartz@rwth-aachen.de (A.R.); amann@lih.rwth-aachen.de (F.A.)Institute of Geomechanics and Underground Technology, RWTH Aachen University, Mies-van-der-Rohe-Straße 1, 52074 Aachen, Germany; foerderer@gut.rwth-aachen.deIndependent Researcher, Altburgstrasse 45, 8105 Regensdorf, Switzerland; sbruelisauer@bluewin.chInstitute of Environmental Engineering (ISA), RWTH Aachen University, Krefelder Str. 299, 52070 Aachen, Germany; linnemann@isa.rwth-aachen.dePhoton Systems Inc., 1512 Industrial Park St., Covina, CA 91722-3417, USA; w.hug@photonsystems.comCorrespondence: cpost@lih.rwth-aachen.de (C.P.); niklas.heyden@rwth-aachen.de (N.H.)2162022720222213412110466823420221562022210620220907202211072022© 2022 by the authors.2022https://creativecommons.org/licenses/by/4.0/Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).

The entire water cycle is contaminated with largely undetected micropollutants, thus jeopardizing wastewater treatment. Currently, monitoring methods that are used by wastewater treatment plants (WWTP) are not able to detect these micropollutants, causing negative effects on aquatic ecosystems and human health. In our case study, we took collective samples around different treatment stages (aeration tank, membrane bioreactor, ozonation) of a WWTP and analyzed them via Deep-UV laser-induced Raman and fluorescence spectroscopy (LIRFS) in combination with a CNN-based AI support. This process allowed us to perform the spectra recognition of selected micropollutants and thus analyze their reliability. The results indicated that the combination of sensitive fluorescence measurements with very specific Raman measurements, supplemented with an artificial intelligence, lead to a high information gain for utilizing it as a monitoring purpose. Laser-induced Raman spectroscopy reaches detections limits of alert pharmaceuticals (carbamazepine, naproxen, tryptophan) in the range of a few µg/L; naproxen is detectable down to 1 × 10−4 mg/g. Furthermore, the monitoring of nitrate after biological treatment using Raman measurements and AI support showed a reliable assignment rate of over 95%. Applying the fluorescence technique seems to be a promising method in observing DOC changes in wastewater, leading to a correlation coefficient of R2 = 0.74 for all samples throughout the purification processes. The results also showed the influence of different extraction points in a cleaning stage; therefore, it would not be sensible to investigate them separately. Nevertheless, the interpretation suffers when many substances interact with one another and influence their optical behavior. In conclusion, the results that are presented in our paper elucidate the use of LIRFS in combination with AI support for online monitoring.

environmental monitoringmicropollutantsdata processingreal-time monitoringwastewater treatment plantDUV Raman/fluorescence spectroscopyartificial intelligenceThis research received no external funding.pmc-status-qastatus0pmc-status-liveyespmc-status-embargonopmc-status-releasedyespmc-prop-open-accessyespmc-prop-olfnopmc-prop-manuscriptnopmc-prop-legally-suppressednopmc-prop-has-pdfyespmc-prop-has-supplementnopmc-prop-pdf-onlynopmc-prop-suppress-copyrightnopmc-prop-is-real-versionnopmc-prop-is-scanned-articlenopmc-prop-preprintnopmc-prop-in-epmcyespmc-license-refCC BY
1. Introduction

One of today’s most pressing global challenges is the availability of a sufficient amount of drinking water of adequate quality. A major threat that is currently posed to surface and groundwater reservoirs is their contamination by micropollutants (pharmaceuticals, pesticides, industrial chemicals, consumer care products). The discharge of harmful substances into the aquatic environment via municipal wastewater treatment plants (WWTP) and combined sewer overflows contributes significantly to the pollution of water bodies [1,2]. Some of these harmful substances appear in µg/L or even ng/L concentrations which overwhelms the sensitivity of many analysis techniques. These so-called ‘emerging micropollutants’ are traces of medical products, plant protection products, biocides and other chemicals that can have detrimental effects on the environment or human health at very low concentrations [3,4]. In 2010, the largest ever EU-wide monitoring survey on WWTP effluents was performed. The obtained results show the presence of 131 target organic compounds in European wastewater effluents, in concentrations ranging from low nanograms to milligrams per liter [5]. Due to the adverse effects on human and environmental health, a legal obligation to reduce and prevent micropollutants is posed by the European Water Framework Directive (WFD) and the environmental quality standards (EQS) for priority substances [3,6,7]. For the substances that are defined in the daughter directive of the WFD the gradual reduction or even cessation of discharges and emissions is to take place [8].

In consequence, specific measures must be implemented to combat water pollution that is caused by relevant pollutants or pollutant groups as soon as environmental quality standards (EQS) are exceeded [9]. These specific measures presuppose that the nature and extent of micropollutants are recognized, for example, in WWTP and their relieving surface water bodies. Unfortunately, most micropollutants are not yet included in the routine monitoring of surface and wastewater [10]. This means that, compared to other anthropogenic contaminants the emerging micropollutants (EMPs) have largely fallen outside the scope of monitoring and worldwide regulations [11]. The occurrence of different organic micropollutants, especially EMPs (even at very low concentrations from point and nonpoint sources) in the natural environment has raised significant concerns about their negative effect on aquatic ecosystems and human health [12,13,14].

Municipal wastewater treatment plants were designed to treat domestic wastewater, not to reduce, for example, pharmaceuticals, pesticides and microplastics. Biologically, WWTP remove protein, fat and carbohydrates, as well as the important plant nutrients carbon, phosphorus and nitrogen from the water [15,16]. It must be considered that today’s conventional wastewater treatment technology (mechanical and biological steps) was designed in the 1970s [5]. This fact underlines to the problem that many substances are not even covered by the normal sum parameter monitoring of a WWTP, thus the chemical inventory of the wastewater is not known, meaning that it cannot be completely purified with the available treatment stages [17]. In fact, most of the conventional WWTP are not designed to eliminate organic compounds at low concentrations, making the treatment processes vulnerable and triggering a dissemination of micropollutants all over the environmental compounds (Figure 1) [18,19].

Information on the degradation of, for example, diclofenac (analgesic) at WWPT ranges from “no effect” to a 60% degradation rate for the different size classes and the different stages of expansion of the WWTP [20,21,22,23,24]. A review paper from Luo et al., 2014, showed that the WWTP removal efficiency for selected micropollutants in 14 countries/regions depicted compound-specific variation in removal, ranging from 12.5 to 100%. It must be noted that the consumption quantities of selected drugs in Germany that are the subject of public debate exceed 1000 t per year in some cases (e.g., for ibuprofen) [9]. In consequence, regarding pharmaceuticals, toxicological studies have shown that they might have direct toxicity towards certain aquatic organisms [25,26,27,28]. The occurrence of micropollutants in the aquatic environment has frequently been associated with several negative effects, including short-term and long-term toxicity, endocrine-disrupting effects and the antibiotic resistance of microorganisms [2,29,30]. Their constant but imperceptible effects can gradually accumulate, finally leading to irreversible changes in both wildlife and human beings [31,32]. The presence of these chemicals in the environment is more concerning considering that they do not appear individually, but as a complex mixture, which could lead to unwanted synergistic effects. The ubiquity of a high number of potentially toxic emerging chemicals in the environment underpins the need to better understand their occurrence, fate and ecological impact [33]. This leads to the recurring demand for online monitoring that follows the concentrations, retention time and chemical interactions between the relevant substances. Therefore, high demands on sensitivity, high throughput by automation and short analysis times are major requirements [34].

Our research tests the possibility of the real-time monitoring of micropollutants in wastewater using laser-induced Raman & fluorescence spectroscopy (LIRFS) in combination with an artificial intelligence approach using convolutional neural networks (CNN). Spectroscopy in general comprises a group of optical techniques that decompose radiation according to the properties of molecules, such as wavelength or energy, producing a spectrum as an excitation response. Radiation in the form of electromagnetic waves maps the specific properties of a medium between the light source and the spectrometer, producing a specific spectrum. The spectrum contains information to confirm atomic and molecular models, determine qualitative compositions, and ascertain the proportion of a substance in a mixture that is based on the spectra intensity [35,36]. In our current research case, we investigate both Raman and fluorescence spectroscopy that was conducted with excitation in the deep-UV below 250 nm on the influent and effluent wastewater of a biological treatment stage and an ozonation stage at wastewater treatment plants. We identified the substances that affect the purification processes in WWTP; influence the installed sum-parameter monitoring (e.g., humic acids); result from purification processes (as a transformation product like bromate); or pass the WWPT without substantial reduction (e.g., pharmaceuticals like diclofenac, carbamazepine), and may lead finally to a degradation of the surface water quality. Our approach aims to identify and quantify the problematic pure substances using LIRFS in combination with a CNN classification model. On every influent and effluent sample fluorescence and Raman measurements were performed and the reliability of the results were statistically evaluated and trained by CNN.

2. Fundamentals2.1. Fluorescence/Raman Spectroscopy

Fluorescence spectroscopy is a method that is based on energy absorption, where a fluorophore molecule is excited into various excited states through the absorption of electromagnetic radiation of a certain wavelength. The radiation is emitted following the transition to the lowest energy state (Kasha rule) and has a longer wavelength (less energy) than the absorbed photon. The wavelength difference between the absorption maximum and the emission maximum is called the Stokes shift [37] (see Figure 2, left). The molecule emits the energy in the form of fluorescent light of a certain wavelength, in order to return from its excited state to the ground state (see Figure 2, right) [38,39].

The process of fluorescence can be described using the simplified Jablonski diagram in Figure 2 (right). In the Jablonski diagram, the energy levels of the ground state (S0) and various excited states (S1, S2, S3), as well as the vibrational levels are shown. A fluorophore molecule can become excited from the ground state (S0) through the absorption of a photon into a higher energy level (S1, S2, S3), indicated by the violet arrow. The excited electron of a fluorophore molecule returns to the lowest vibration state of the first excited state (S1) due to internal conversion (blue arrows). From the lowest vibration level, the electron falls to the ground state by emitting the energy due to fluorescence (green arrow).

Another optical method is Raman spectroscopy. The Raman effect can be described as a two-photon process. Raman scattering is observed when monochromatic light hits a molecule and deforms or polarizes the electron shell, creating a short-lived so-called ‘virtual state’ of the electron. The energy absorption of the incident photon is called anti-Stokes scattering, whereas the energy transfer to an emitting photon is called Stokes scattering [41,42].

As we mentioned in our previous paper [43], deep-UV spectroscopy offers the opportunity to analyze Raman and fluorescence emissions separately by using a nearly fluorescence-free wavelength range.

This was first demonstrated by Asher et al. [44]. As Liu [45] summarized, a wavelength range was discovered in which the interfering fluorescence can be separated from the Raman signals.

Photon Systems Inc. [46] showed a spectral bandwidth over 3000 1/cm in which no interference from fluorescence was observed by DUV-Raman excitation at a wavelength that was below λ = 250 nm. This effect is well shown by Photon Systems on the Raman and fluorescence spectra of crude oil that is excited by different wavelengths [47,48].

The fluorescence behavior and intensity can be significantly influenced by different chemical and physical factors, called quenching. Quenching in general refers to all processes that decrease the fluorescence intensity, such as reactions in the excited state, the transfer of energy, molecular collisions and complexation reactions. The term collision quenching refers to intermolecular interactions, e.g., in the form of shocks, by which the excited fluorophore returns to ground state after colliding with a quencher [39]. The mechanism of fluorescence quenching depends on the colliding molecule pair, especially on the substances which act as quenchers (in particular, oxygen, halogens, amines and molecules with electron deficiency). In addition to quenching, the internal filter effect and metal ions in the water influence the fluorescence behavior.

The fluorescence intensity is furthermore greatly influenced by temperature, whereby the correlation between these factors is anti-proportional. In the event of a temperature rise, the probability that an excited electron is caused by radiation-free decay increases. How great the effect on the fluorescence intensity is depends on the composition and size of the fluorophore [49].

The complexity of the largely unexplored behavior between the individual relationships and thus their influence on fluorescence is the subject of numerous works in the literature [50,51,52].

2.2. Wastewater Treatment and Monitoring

Wastewater treatment developed rapidly in the last decades due to ongoing challenges in wastewater composition. All the purification processes are undergoing a process of optimization, so this brief overview emphasizes a bundle of applied process combinations but cannot guarantee completeness.

The aim of the first treatment stage is to remove the undissolved components from the wastewater. In this process, coarse materials are retained by screens, sediments are forced to sediment by lowering the flow velocity, and floating light materials are removed by skimmer devices. A further, detailed explanation is given in [53,54,55]. In Gupta et al., 2012 [56], a good illustration of primary, secondary and tertiary treatments and water recycling technologies is shown.

Secondary water treatment includes biological routes for the removal of soluble and insoluble pollutants by microbes [57]. The microbes, usually bacterial and fungal strains, convert the organic matter in an aerobe process (nitrification) and later in an anaerobe process (denitrification) into methane and inorganic substances, such as carbon dioxide and ammonium [58]. Many microorganisms form colonies which become visible as sludge flocs. These sludge flocs settle at the bottom of the subsequent secondary settling pond or clarifier and are either returned to the activated sludge process or pumped into the primary settling pond for sludge disposal (see Figure 3). The biological treatment includes the aerobic and anaerobic digestion of wastewater. The aerobic process is effective for the removal of biological oxygen demand (BOD), chemical oxygen demand (COD), dissolved and suspended organics, volatile organics, organically bound nitrogen and ammonia, phosphates etc. The concentration of biodegradable organics can be reduced by up to 90% [56]. If free-dissolved oxygen is not available in the wastewater, anaerobic decomposition occurs. Anaerobic and facultative bacteria convert the complex organic matter into simpler organic compounds that are based on nitrogen, carbon and sulphur [58]. The important gases that are involved in this process are nitrogen, ammonia, hydrogen sulphide and methane. In our case study, the influent and effluent of a membrane bioreactors (MBR) were examined, which belongs to the biological treatment stage. MBR is a membrane-filtration system that can be integrated directly into the aeration tank (Figure 3). Micro- and ultrafiltration membranes are used for this purpose [59] which separate the activated sludge, thus eliminating the need for the secondary clarifier.

The third purification stage is important to reduce different nutrients, such as phosphate and nitrogen, but especially for disinfection purposes. WWTP design often varies due to different boundary conditions, so that the elimination of phosphate in some WWTP belongs to the biological purification stage [60]. Various possibilities of third water treatment processes provide a final stage to meet specific requirements around the safe discharge of water. In our case, ozone was used as an oxidant. Although ozonation has proven to be an effective means of eliminating trace contaminants, undesired transformation products (TPs) are also formed in this process [28]. In this context, the formation of bromate is by far the most critical problem. The studied ozonation stage is controlled in a minimum operation with a volume proportional dosing of 2.0 mg O3/L. For the evaluation, samples collected before and after ozonation were compared over a period of two weeks.

Overview flow scheme of the WWTP Aachen-Soers. Numbers represent the steps of the treatment process: (1) fine and coarse screen, (2) grit removal, (3) primary clarifier, (4) anaerobic denitrification, (5) aerobic nitrification, (6) secondary sedimentation, (7) ozonation, (8) second nitrification, (9) sand filter. (EF) effluent into surface water [61] (adapted with permission from [61]. Copyright 2022, Elsevier).

As described above, the most important removal pathways of organic compounds during wastewater treatment are biotransformation/biodegradation, abiotic removal by adsorption onto the sludge and stripping by aeration (volatilization) [62]. To track and review these interacting processes, sum parameter monitoring is utilized. Sum parameter bundling represents a summarized description of certain effects and substance parameters. Due to the large number of organic compounds in (waste)water, the substances are often not considered as individual substances but are grouped together according to their similar properties or effects [63]. This offers the advantages of easy, standardized and affordable measurements. Consequently, they belong to the most frequently measured parameters [27]. However, they do not provide any detailed information on the presence of toxic or emerging contaminants.

The BOD (biological or biochemical oxygen demand) indicates the amount of oxygen that is consumed by microorganisms for the biological, aerobic degradation of dissolved organic compounds in wastewater within a certain time. It is usually given in mg O2/L for a period of 5 days (BOD5) [64]. The source of BOD is readily degradable organic carbon and ammonia (NH3). These are common constituents or metabolic byproducts of plants and animal wastes and human activities [65]. The standard method for COD estimation is the so-called ‘Dichromate’ method [66,67] which is characterized by the acidification of the sample with sulphuric acid and the addition of silver sulphate. Due to the application of hazardous chemicals and an analysis time of 2 h, the method is not suitable for online use [68].

As a result of the implementation of the EU Industrial Emission Directive (IED), the total organic carbon (TOC) parameter is being introduced broadly into the monitoring of wastewater in various industries that reflects the organic load in a sample and is divided into dissolved organic carbon (DOC), undissolved organic carbon POC (particulate organic carbon), and volatile organic carbon (VOC) [69]. There is a slight preference of TOC, compared to the determination of COD, because TOC is subject to uniform regulation across Europe by the [70] standard, it is not hazardous, and it includes only organic carbon compounds. Generally, the determination of TOC is achieved by thermal or wet chemical oxidation so that CO2 is formed, which is subsequently measured by an NDIR (non-dispersive infrared) detector. The distinction between DOC and POC lies at a particle diameter of 0.45 μm [71]. On average, DOC represents 87% of TOC [72]. The DOC in wastewater is mainly composed of easily degradable substances, such as carbohydrates, proteins, fats and amino acids, as well as compounds that are difficult to degrade, such as cellulose and humic substances. Hence, particles which contain organic carbon should be considered during the measurement [70]. DOC content is reduced by up to 90% in WWTP plants. In Germany, however, there is no limit value for DOC in drinking water or wastewater, while Swiss drinking water may contain a maximum of 2 mg DOC/L [64].

Process analyzers, such as gas chromatographs with simple detectors [73,74,75], TOC-analyzers, spectroscopic probes and many others are very common in the process environment and have already been implemented successfully in (petro-)chemical plants or wastewater treatment plants for many years [76]. These techniques (e.g., GC, HPLC and IC systems) mainly use mass spectrometric detection which does not allow truly continuous monitoring. The spectral absorption coefficient (SAC254) provides information on the load of dissolved organic matter in the water. The coefficient is determined using a spectrophotometer at a wavelength of 254 nm and gives the absorption of the dissolved substances in the sample [77]. Modern optical sensors measure the turbidity according to EPA 180.1 [78] and ISO 7027 [79]. Other measuring probes apply UV-vis spectroscopy over the entire absorption spectrum from 190 to 750 nm. These sensors enable in-situ measurements during operation and real-time adjustments of the treatment process. However, the detection limit in waters with a complex composition and the difficulty of detecting the UV-Vis spectra of some pollutants in the water, such as suspended solids, dissolved inorganic substances and pathogenic microorganisms remain challenging [80]. Monitoring tools that interact continuously and automatically are not yet established. As [81] summarized: “Analytics 4.0 was proclaimed before it became a reality, and at least for the traditional analytical laboratory describes a vision rather than the actual status. Standardized protocols and commercial, unified solutions are required”.

Raman- and fluorescence laser spectrometry offers the potential to replace or supplement these methods and to establish faster and more efficient real-time monitoring as a new standard. This could act as an early warning system to detect targeted substances and allow rapid intervention.

3. Equipment3.1. Raman/Fluorescence Detection System

For this study, we utilize a commercially available Photon Systems PL200 DUV spectrometer [47]. The system uses the benefits of deep-UV spectroscopy by using a NeCu transverse excited hollow-cathode gas laser, which leads to an interference-free spectral range without fluorescence and the auto-luminescence obscuration of Raman signals, and vice versa [45,47,48]. Most of the common devices that are used in spectroscopy are relatively large, heavy, and require significant cooling power. The PL 200 mitigates these limitations by utilizing a compact, low power, air-cooled hollow-cathode laser. The system is discussed in more detail in [43] and only the main components will be summarized. The laser emits light with a wavelength λ = 248.6 nm in 40 µs pulses with an average energy of 3.5 µJ. The device is designed to carry out fluorescence and Raman measurements which is enabled by the low wavelength, allowing a clear separation of the Raman shift signals and the autofluorescence of many natural materials.

Figure 4 shows a detailed illustration of the measurement setup. The incident laser light is focused by an array of mirrors where the aperture and light intensity are controlled by two iris diaphragms. An interference filter ensures a monochromatic light. Only one wavelength of 248.6 nm is transmitted through an additional built-in interference filter. The light that is reflected back from the sample reaches the laser-blocking lens via an achromatic objective. This lens almost reflects the Rayleigh radiation that is emitted by the sample and is transparent to the incoming Raman radiation. The sample is measured inside a quartz glass cuvette, a movable flow cell or placed onto a rotary table. The emitted fluorescence signal is then edge-filtered and spectrally separated by a monochromator. The light enters the monochromator through an entrance slit, the size of which can be adjusted. A larger slit decreases the spectral resolution. At the same time, enlarging the slit leads to an improved limit of detection (LOD). Since these two properties are linked ant proportionally, a trade-off exists between the resolution and LOD. We choose a relatively large slit size of 150 µm, in order to achieve better detection limits, since the LOD is much more important when trace amounts of substances need to be identified. After passing through the entrance slit, a 200 mm focal length Czerny–Turner spectrograph splits the light into the individual wavelengths. The fluorescence signal is acquired by selecting a 300 ln/mm grating, which yields a resolution of 1.2 nm. The 3600 ln/mm grating can be chosen, in order to investigate the sample’s Raman shift, which is investigated in a previous publication [43].

The resulting light spectrum is detected by a three-stage thermo-electrically cooled, back-thinned, back-illuminated CCD array with a resolution of 2048 × 128 pixels. The intensities are recorded with 16-Bit resolution, which means a maximum intensity of 65,000 counts can be recorded.

The emission is based on a back-scattering configuration (180° degree) when using the cuvette from Hellma® or Starna CellTM. For the detection of solid samples, such as microplastics or powders, a rotary table can be used, resulting in a 180s degree scattering configuration.

The spectrometer can be calibrated quickly and easily via software (Figure 5, left). The Raman grating (3600 ln/mm) uses Acetonitrile, a NIST-Standard (National Institute of Standards and Technology) for calibration. The calibration is based on two known peaks at 2250/3940 1/cm (compare Figure 5a) and can be verified with pure water (Photon Systems Inc. 2020). The fluorescence calibration is achieved by using a GaN-sample (gallium-nitride), showing one specific peak at 363 nm (Figure 5b). The calibration can be verified by measuring the amount of PHE (Phenylalanine) [82].

4. Data Processing4.1. Data Preparation

All the data were processed and analyzed using the Spectrum Analyzer© software. The software is also used to control and calibrate the two holographic gratings. In addition, the software supports the use of CNN classifiers, which we trained using our measured spectra.

For the software optimization, each measurement is repeated 10 times, in order to improve the signal-to-noise ratio. The recorded spectra are processed to enable better comparability and to optimize the quality of the spectra. The fluorescence and Raman spectra are handled with the same mathematical algorithms. All spectra are smoothed using a Whittaker–Henderson algorithm. Additionally, sharp and intensive peaks, which can be caused by cosmic rays are removed (despiked). The intensities are then normalized regarding the pulse number and laser energy due to variations in the laser energy.

For the statistical analysis, each Raman or fluorescence measurement series, consisting of 10 individual measurements to evaluate the precision (and photodegradation or photobleaching) is preceded by a calibration process, as described before in Section 3.1. Dilution series with decreasing substance concentrations in the solution focus on the accuracy of the results in terms of trueness, and lead to the estimation of the limit of detection for each substance. Depending on the fluorescent properties, some of the substances are stronger diluted, meaning that they show more steps regarding their dilution (e.g., naproxen or tryptophan). The standard deviation (σ) and coefficient of variation (CV) describe the characteristics of the single spectrum and result in a correlation curve, if the results so indicate. σ=(∑x−x′)2n CV=(∑x−x′)2nx′ with x = intensity of signal, x′ = averaged intensity of signal, n = number of values.

The further processing refers to the fluorescence measurements, whereas Raman data processing was described in detail in a previous paper [43]. The averaged fluorescence spectrum of the individual concentration is used, from which the spectrum of the solvent has already been subtracted if the substance spectrum has a higher signal intensity than the solvent. In the other case, to prevent negative values for the resulting spectrum, we normalize the solvent spectrum to its initial, sharp peak (Figure 6). This procedure artificially increases the substance signal, in order to reduce masking by the solvent. If this procedure is necessary, a quantitative view on the date is no longer possible, regarding the limits of detection or the evaluation of concentrations.

In the further course, the maximum signal intensity of the different concentrated solutions is determined. The corresponding intensities are plotted against the concentration and a calibration curve is generated. From this, the mathematical detection limit of fluorescence spectroscopy for the respective substance can be determined graphically, as well as the accuracy of the measurements. For this purpose, the values of the calibration curve are used as reference values and the relative deviations of the measured spectra to this curve are calculated. If a linear regression results, the coefficient of determination (R2) can be used to estimate the accuracy. If the calibration curve does not yield a linear regression, the relative deviation is determined with Equation (3) [83]. rel. σ=measured value−true valuetrue value

For the identification of the measured substance, the spectrum of the pure substance as a solid, gas or liquid is measured and stored by the database. References for Raman spectra can also be found in the literature or in open-source software [84]. A simple example for the fluorescence spectra of a dilution series of four concentrations of carbamazepine, diluted in ethanol, and measured in a quartz cuvette is shown in Figure 6.

Fluorescence spectra of four carbamazepine concentrations, measured in a quartz cuvette and diluted in ethanol. The spectrum of the solvent ethanol (dotted line black), measured in a quartz cuvette and the empty cuvette itself (dotted line grey) are shown but not yet subtracted. Settings: pulse number 50; pulse frequency 40; slit size 150 µm; grating 300 ln/mm; focal length 20 mm (raw data from [85]).

Figure 6 shows the dominance of the solvent ethanol and its influence on the carbamazepine signals. The highly concentrated 10 mg/g carbamazepine signal (purple line) is most attenuated by the solvent and the other substance spectra become increasingly similar to the ethanol spectrum as their concentration decreases (with the exception of 0.01 mg/g, green line). On the other hand, the signal of the quartz cuvette (dotted line grey) offers a very low intensity so that it can be neglected for further fluorescence measurements. The before-described procedure to reduce the influence of the solvent shows Figure 7, where the ethanol spectrum was subtracted and the averaged carbamazepine spectra were scaled and therefore artificially strengthened. As a result, the highly concentrated solution shows the highest intensity of more than 60,000 AU.

4.2. Substance Classification

In order to use the gathered spectra for monitoring purposes, the recorded substances must be identified. Various techniques for spectrum classification have been developed in the past few years. However, there is no wholly developed study applying these techniques to assist wastewater treatment operation [86]. The complex, nonlinear nature of spectroscopic data makes their analysis and classification challenging. A wide range of strategies for the analysis of spectral data exist, ranging from visual inspection by an expert through database matching, up to sophisticated machine learning models. In the analysis of complex mixtures, especially, the visual identification or correlational database-matching of spectra are often not applicable, since a superposition of all individual molecules’ fluorescence signal is recorded. Furthermore, manual classification is not a suitable workflow for the automated identification of substances. To reduce subjectivity and enable automated workflows we utilize a convolutional neural network (CNN) to classify the spectra. CNNs have proven to be suitable for the classification of spectral data and they need little data pre-processing, such as a baseline removal.

The architecture of our classifier is the same as the Raman spectra classifier that is introduced by [87] and which is also employed in our previous publication on Raman spectroscopy [43]. The only change in the network architecture is the number of output classes. Our investigations focused on a selection of 23 critical substances. The data pre-processing consists of resampling of the wavelengths to values ranging from 0 to 4000 and normalization of each individual spectrum’s intensities to a range between 0 and 1. A signal-to-noise ratio threshold of 3 was also implemented to avoid fitting the model to very noisy spectra. A total of 1700 spectra were used for training; 566 spectra were used as a test dataset. Convergence was achieved after 50 epochs of training using the Adam optimizer and categorical cross-entropy as the objective.

The resulting classifier has an accuracy of over 95% across all substances on unseen spectra. Performance across all classes was close to a 100% true positive rate, except for NaOH (sodium hydroxide) which is consistently misclassified, as well as tryptophan. Furthermore, the spectra of polymer suspensions can be subject to misclassification. This section may be divided by subheadings and it should provide a concise and precise description of the experimental results, their interpretation, and the experimental conclusions that can be drawn.

5. Methodology and Targets

The article at hand offers a methodical case study about a laser-induced Raman and fluorescence spectrometer (LIRFS) that is used in a wastewater treatment plant for monitoring purposes. The application in such a heterogenic and challenging environment is critical for an analytical device that needs enough sensitivity for detections limits down to a few µg/L, but also a robustness, so that the results are not affected by the presence of a variety of other chemicals and substances. The first step to test the suitability of the LIRF spectroscope was the Raman and fluorescence measurement of all the pure substances (as a powder or solid) and the substances in a solution varying in concentration, in order to evaluate reliable identification down to a concentration level, where a lower rate than 3/1 between signal and noise was observed. This was tested on different kinds of problematic micropollutants, such as pharmaceuticals, listed in the (Table 1) below, on two-amino acid tryptophan and tyrosine and on byproducts of the wastewater cleaning process (nitrate, nitrite, bromate, bromide).

In order to avoid the well-known problem of overlapping of the Raman spectrum by the generally more intense fluorescence spectrum, an excitation wavelength in the deep ultraviolet radiation range was selected. Fluorescence spectra that always emit at the same wavelength regardless of the excitation wavelength appear predominantly in wavelength ranges that are above 270 nm. In contrast, the Raman spectra show a dependence of the excitation wavenumber that is used. If the substrate is irradiated with deep ultraviolet light, the Raman spectra result in a fluorescence-free wavelength range that is below 270 nm [89]. Due to this fact, we could measure the Raman spectrum and the fluorescence spectrum of every substance in short succession only by changing the diffraction grating of the device, which takes a few seconds.

As mentioned before, we first measured the pure solids, e.g., powder, to obtain the typical substance spectrum for recognition in the database, and then set up a dilution series to evaluate the sensitivity of the spectrometer. Depending on the substance and its solubility, the solutions are produced with deionized pure water (Millipore®, purity ≥ 99%, Merck KGaA, Darmstadt, Germany), ethanol (C2H6O; purity ≥ 96%, Carl Roth GmbH & Co. KG, Karlsruhe, Germany) or as standardized solutions for ion chromatography. Therefore, the optical properties of the solvent and the used cuvette (quartz glass) must be considered, or their signal must be subtracted from the resulting spectrum. Pre-tests showed a light optical effect of the quartz cuvette for the fluorescence measurements (maximal 600 AU), whereas there is one regarding the Raman measurements (Figure 8). In order to bypass this obstacle, the heterogeneous WWTP samples regarding the byproducts were measured in a flow cell that was constructed with sapphire glass, which had no effect on either fluorescence or Raman scattering.

Regarding WWTP sampling and sample preparation, all the wastewater samples were collected over a maximum of 24 h and mixed without any other treatment but keeping them cool and dark in the refrigerator. Figure 9 shows the extraction points for the different targets coming from two different WWTP.

Regarding online wastewater monitoring, another issue is the evaluation of the assumed correlation between the sum parameter DOC and the fluorescence intensity in the biological treatment stage [91,92,93]. The evaluation of these in a literature-published hypothesis revealed a correlation between the tryptophan-like peak and DOC, which was another focus of the current investigation. According to [94], tryptophan is especially useful as a reporter for conformation changes in proteins, since several protein changes can result in predictable changes in tryptophan fluorescence in a sample [94]. On the other hand, [95] found strong correlations between tryptophan and the colony counts of a variety of indicator species, including E. coli. bacteria, which are extremely important for wastewater disinfection, especially since online monitoring is not yet state of the art.

6. Results

The results of this study originate from different field and laboratory campaigns. All the samplings and experiments were performed under standardized conditions so that comparability could be assumed. Deviating framework conditions are explicitly mentioned and are justified in the respective objective of the experiment. The selected results underline the possibility to apply the evaluated spectrometer for online monitoring purposes, but they also emphasis difficulties and obstacles in operation.

We first examined the capability of fluorescence intensity measurements to correlate with organic carbon concentrations. According to [93], fluorescence peaks C in water can be used to estimate DOC, where peak C correlates with the content of dissolved organic carbon in the water. Our results could confirm these observations and transfer them to treated wastewater (Figure 10). The fluorescence analysis of wastewater samples from two different WWTPs after passage through the membrane bioreactor are shown in Figure 9 [96]. The marked spectra correspond to the DOC concentrations that are estimated in the collaborating laboratory via thermal oxidation and NIR detection DIN EN 1484. There is a clear tendency that the higher the DOC concentration, the higher the peak intensity (except for sample 163). The coefficient of variation from these seven samples ranges from 0.5 to 4.7.

In our measurements, the coefficient of determination R2 is 0.85 (Figure 11), referring to seven effluent samples of two different WWTPs that are equipped with a membrane bioreactor. This corresponds to the results from the literature, e.g., from [97] of R2 = 0.8 and 0.81 from [98]. The correlation between the fluorescence intensity in all our measurements and the DOC concentration determined in the laboratory showed that a differentiated consideration of the extraction point was useful. The correlation coefficient of all the samples throughout the purification processes was 0.74 [96]. Variations show up more clearly in lower concentration ranges than in higher ones, but the results suggest that the correlations allow a rough estimation of the concentrations of DOC, and every determination of DOC in the lab could be reproduced by the spectrometer. In particular, anomalies—such as a sudden increase in DOC—can be confidentially detected that allow early warning and a rapid response, for example, the application of ozone.

With the aid of the dilution series, the detection limit for the amino acids tyrosine and tryptophan could be determined. Tyrosine proves to be detectable up to 0.05 mg/L, but for tryptophan the spectrometer shows a much greater sensitivity down to 0.001 mg/L (Figure 12b). At lower concentrations, the intensities continue to decrease, and the peaks are strongly affected by solvent and background noise.

For comparison the results of tyrosine is shown in Figure 13.

No correlation was found between the tyrosine-like peak and DOC in wastewater, which is probably because tyrosine fluoresces at an excitation wavelength that is lower than 248.6 nm and the presence of quenchers in the wastewater reduces the fluorescence properties of tyrosine. Regarding the correlation between the tryptophan-like peak in the fluorescence measurements and DOC, ref. [99] showed a high correlation between the (tryptophan-like) peak and DOC concentration (R2 > 0.85) and concluded that this peak can be used for monitoring DOC in untreated and treated water. Our results relativize this observation, depending on our excitation wavelength, and suggest a more differentiated approach regarding treated wastewater. Overall, a medium correlation can be demonstrated between the DOC concentration and the peak intensities of T2 (Figure 14).

Considering all samples, the coefficient of determination is R2 = 0.75. This result is largely in line with the literature values [94,95]. Differentiating the extraction points shows that the correlation is significantly higher for the effluent samples (R2 = 0.95) than for the influent samples (R2 = 0.19). This leads to the conclusion that the correlation is higher for water with a low DOC concentration (≤8.3 mg/L in our samples) than for samples with a higher DOC content (≥37 mg/L in our samples). This may also be related to the presence of other interfering substances in the water which are removed in the aeration tank, but strengthens the previous results of the membrane bioreactor.

Analyzing the spectra of heterogeneous WWTP samples is extremely challenging due to effects that are caused by coexisting substances and their interactions. Correlations do not appear to be consistently linear, and the spectra interact with substances that remain in the treated wastewater. Figure 15 shows six influent and six effluent samples from the biological treatment stage. Samples with a similar color were taken on the same day (influent samples have a solid line; effluent samples’ lines are dotted) at the entry and at the output of the aeration tank.

The reduction rates of tryptophan peak T2 through purification vary for the sample pairs from a maximum of 80.0% (sample number 634 and 635) to a minimum of 32.2% (sample number 706 and 707) (Figure 15). Effluent samples number 703 and 707 are characterized by a low purification efficiency and lead to high contents of higher-ring hydrocarbons in the wavelength of DOC with an emitting wavelength of around 440 nm. The corresponding reduction in the DOC rates shows a similar trend, but the amounts are not transferable. The DOC reduction in sample pair number 634 and 635 amounts to 90.6%, and for sample number 706 and 707 the DOC content was reduced by 79.3%, which differs from the decrease in tryptophan content in the same sample pair. The variation coefficient for the calculated DOC concentrations referring to the correlation curve in Figure 14 leads to values of more than 200% in some cases.

Regarding the pharmaceutical micropollutants, we demonstrate relevant results showing the ability of the spectrometer to detect specific substances, and its suitability for monitoring purposes. The spectra of metformin hydrochloride as a solid shows the trend of photodegradation after 10 measurements (Figure 16).

The metformin hydrochloride dilution series do not show very meaningful results in the fluorescence and Raman measurements, but low signals down to 0.01 g/L are detectable. The metformin hydrochloride signals are strongly affected by the water emissions, but for the Raman measurements a clear peak at 940 1/cm was recognized that was not superimposed by the Millipore® spectrum. Carbamazepine by far shows stronger fluorescence properties. Upon measuring the solid powder, the CCD detector is completely oversaturated so that the full spectrum is not visible. Dilution was carried out with 96% ethanol. Figure 17 shows the fluorescence and Raman spectra of the diluted carbamazepine.

Remarkable in both measurements is the influence of ethanol on the carbamazepine spectra in such a way that the peaks of ethanol drag along the spectra of carbamazepine, especially in the low concentrations. Acetaminophen shows an intensive Raman peak at 1607 1/cm, measured with 2000 pulses. After the correction of the rolling baseline, the variation coefficient remains relatively high (12.8%) due to slight photodegradation effects during 10 repetitions. The fluorescence measurements with 100 pulses could confirm high variations but without a clear trend. The fluorescence dilution series showed good differentiation between the concentration steps and were detectable down to 0.1 mg/g.

Hydrochlorothiazide powder is specifically recognizable by its fluorescence signal with a strong peak at 373 nm (measured with five pulses) with an intensity of more than 45,000 AU and a good precision that is represented by a variation coefficient of 1.9%. However, the dilution series revealed no correlation between the concentration and signal intensity, so that the construction of the calibration curve was not useful.

On the contrary, naproxen is detectable in solution with ethanol in the fluorescence spectrum down to 1 × 10−4 mg/g (Figure 18). Theoretically, this enables its application in the WWTP monitoring. Even the Raman measurements show that two characteristic peaks at 1373 1/cm and 1616 1/cm recognize naproxen specifically.

Regarding the Raman measurements (Figure 19) for the nitrate dilution series, a pulse number of 500 was used for the solutions containing 50 mg/L, 25 mg/L, 10 mg/L and 5 mg/L NO3−. On the other hand, for the solution with 1 mg of NO3/L, a pulse number of 1000 was used. The corresponding spectrum was then scaled by a factor of 0.5 to establish comparability with the other spectra in the dilution series. This procedure is generally acceptable because there is a linear relationship between the applied pulse number and the measured intensity. However, problems can occur in the range of high concentrations due to nonlinear behavior in the form of strong absorption and reemitting.

The applied function of the calibration curve (Figure 20) is concave, shows a very good correlation with the measured values with a coefficient of determination of 0.99 and can predict the measured peak intensities accordingly well. The relative deviation of the data around the arithmetic mean ranges from 2.31% to 3.97% with an average variation coefficient of 3.03%, indicating a high level of precision.

Nitrate is not detectable in the fluorescence measurements but is clearly and specifically detectable down to 10 mg/L of concentration (Figure 19b, green line) in the Raman measurements. This is not convincing for the WWTP application for the effluent water, even though values of this concentration sometimes exist. Further, of course, an anomaly warning in case of unexpected loads entering the wastewater treatment plant is possible as pre-studies from the food industry have indicated. The same behavior of nitrate is found for nitrite showing a lacking fluorescence emission; even the Raman intensities are little lower than for the nitrate spectra (Figure 21).

Regarding the Raman dilution series of nitrite, a reduction in the peak intensity is shown with each dilution step, without exception. Up to a nitrite concentration of 10 mg/L a clear peak is formed in the spectra. At a nitrite concentration of 5 mg/L only a broadened base of higher intensity can be seen. The applied function of the calibration curve is convex, shows a very good correlation with the measured values with a coefficient of determination of 0.98 and can predict the measured peak intensities accordingly well with an average variation coefficient from 2.27.

A more problematic substance to detect is bromide, which was also measured in the flow cell. Since the measured spectra of the dilution series do not show any peak that could originate from the bromide. Furthermore, the measured spectra differ only marginally from each other, despite strongly different bromide concentrations. The expected Raman wavenumber range of the bromide anion agrees with the statement of [100] that most anions of inorganic salts have peaks in the wavenumber range of 500–1100 1/cm. Since these peaks occur in the region of the enveloping curve of the water, there is a possibility that the bromide peak is superimposed by the intense signal of the water. Another possible explanation is provided by the basic operation of Raman spectroscopy, since the Raman signals measured are primarily due to vibrations of the electron bonds between two atoms of a molecule. However, in the bromide standard that was used, the potassium bromide (KBr) was dissolved in water and therefore exists in its dissociated form as Br- anions and K+ cations. Consequently, no electron bond exists between the bromide and another atom that could be excited to vibrate.

The bromate dilution series was carried out analogously (Figure 22) to the bromide dilution series. Here, too, a total of six solutions with different concentrations were prepared and sampled as standard with a pulse number of 500. The two solutions with the lowest concentrations were analyzed with a pulse number of 1000 for visibility reasons.

The spectra of the bromate dilution series show similarity to the bromide series in that there are no specific peaks in the spectrum that cannot be found in the spectrum of the solely Millipore® water. Figure 23 shows the reference spectrum of solid potassium bromate (KBrO3). The spectrum shows that no Raman-active signals exist above a wavenumber of 900 1/cm. The peak with the highest intensity occurs at a wave number of approx. 800 1/cm.

The Raman signals occurring in the wavenumber range between 150 1/cm and 200 1/cm are attributed by [101] to the vibration of the crystal lattice of the solids and therefore do not occur when the bromate is presented in its dissociated form. The two most intense Raman peaks of the bromate anion represent the ν1 and ν4 vibrations, which occur at wavenumbers of about 800 1/cm and 420 1/cm, respectively. Thus, it can be stated that the bromate anion, in contrast to the bromide anion has Raman-active vibrations and can in principle be detected with the applied technique, but the peaks of the vibrations are not visible in the measured spectra. The most likely explanation is that the peaks of the bromate anion are superimposed by the enveloping curve of the water.

Summarizing the results of all measurements, Table 2 gives an overview of the performed analyses and the ability of the spectrometer that was used to detect the different substances.

7. Discussion and Conclusions

The presented results are manifold. It should be noted that both measuring techniques that are included in one device have specific strengths and weaknesses. The fluorescence technique is very suitable for observing DOC changes in wastewater. This technique can provide a good service to wastewater treatment plants by offering online monitoring that checks the proper operation of the purification process and sending a warning when it is threatened by sudden heavy pollutant loads. This could be proven by the investigation results in the aeration tank, the membrane bioreactor and at the ozonation. The results prove that changes in concentration ranges of a few mg/L DOC could be observed, and these correspond to the results of the cooperating laboratory. Generally, it could be confirmed that high concentrations of various co-existing substances, as in wastewater, lead to the underestimation of substance quantities (by fluorescence), probably due to the inner filter-effect and quenching. Another weak point becomes apparent in the identification of individual substances. For example, fluorescence can detect changes very sensitively, and it can indicate which substance transformations have taken place, but this does not lead to the proper identification of substances. Here, only the observation of substance groups can be used to determine these substances based on their typical wavelengths.

The Raman measurements impressively demonstrated that selected pharmaceutical substances (carbamazepine, naproxen, tryptophan) could be detected down to a few µg/L range—this is a concentration range that is within the range of real application in a sewage treatment plant. However, some of these micropollutants indicated that due to their low Raman activity, or even superposition by the water/solvent signal, the detection limits were possible only up to the mg/L range or even the g/L (metformin hydrochloride) range, making it impossible for use in wastewater monitoring. Furthermore, the monoatomic anions (e.g., bromide) of dissolved salts are a problem in Raman spectroscopy since they do not provide a direct Raman signal. Thus, it must be concluded that the monitoring of these substances using Raman spectroscopy is not practical. In the instance of nitrate, the detection limit seemed somewhat weak (roundabout 10 mg/L); however, the Raman detection of nitrate after biological purification in a wastewater sample was successful, as [43] shown. This exact identification is possible due to the AI support which could reach a successful assignment rate of over 95%. Once a substance is trained by the CNN and stored in the database, its identification becomes much easier and more reasonable. For this reason, LIRF spectroscopy seems to be a promising tool for online monitoring, and even device-specific adjustments like the variation of the excitation wavelength could be particularly helpful for the application case of a wastewater treatment plant.

Acknowledgments

Thanks to Jens Fischer for the DOC, TOC and pharmaceutical measurements and Reiner Gschwendtner for anions detection in the environmental samples.

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author Contributions

Conceptualization, F.A., C.P., N.H. and V.L.; methodology, S.B., N.H., A.F. and C.P.; software, S.B. and N.H.; validation, C.P., N.H. and A.R.; formal analysis, S.B., N.H. and A.R.; investigation, N.H., A.R., A.F. and V.L.; resources, F.A.,V.L. and W.H.; data curation, A.F., N.H. and A.R.; writing—original draft preparation, C.P., A.R. and N.H.; writing—review and editing, F.A., W.H. and V.L.; visualization, N.H. and A.R.; supervision, F.A., C.P. and V.L.; project administration, C.P.; All authors have read and agreed to the published version of the manuscript.

Institutional Review Board Statement

Not applicable.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References SacherF. Kümmel VThomaA. FuchsS. KaiserM. LambertB. UllrichA. Analytik von prioritären Stoffen in Abwasserproben—Eine wichtige Voraussetzung für eine Bestandsaufnahme in deutschen KläranlagenMitt. Umweltchem. Ökotoxikol.201835658 LuoY. GuoW. NgoH.H. NghiemL.D. Ibney HaiF. ZhangJ. LiangS. WangX.C. A review on the occurrence of micropollutants in the aquatic environment and their fate and removal during wastewater treatmentSci. Total Environ.2014473–47461964110.1016/j.scitotenv.2013.12.06524394371 Umweltbundesamt Recommendations for reducing micropollutants in watersGerman Environment Agency Section II2.1 General Aspects of Water and SoilUmweltbundesamtDessau, Germany2018 RasheedT. BilalM. NabeelF. AdeelM. IqbalH.M. Environmentally-related contaminants of high concern: Potential sources and analytical modalities for detection, quantification, and treatmentEnviron. Int.2019122526610.1016/j.envint.2018.11.03830503315 LoosR. Negrão De CarvalhoR. ComeroS. CondutoA.D. LettieriG.M.T. LocoroG. ParacchiniB. TavazziS. GawlikB. BlahaL. EU Wide Monitoring Survey on Waste Water Treatment Plant EffluentsEUR 25563 EN, JRC76400Publications Office of the European UnionLuxembourg2012978-92-79-26784-0WFD Directive 2000/60/EC of the European Parliament and of the Council of 23 October 2000 Establishing a Framework for Community Action in the Field of Water PolicyAvailable online: https://eur-lex.europa.eu/resource.html?uri=cellar:5c835afb-2ec6-4577-bdf8-756d3d694eeb.0004.02/DOC_1&format=PDF(accessed on 22 April 2022)Directive 2008/105/EC Environmental Quality Standards (EQS) Applicable to Surface WaterAvailable online: https://www.legislation.gov.uk/eudr/2008/105/contents(accessed on 22 April 2022)LANUV NRW Untersuchungen zum Eintrag und zur Elimination Gefährlicher Stoffe in Kläranlagen. 2006, Teil 2Available online: https://scholar.google.com/scholar?hl=en&as_sdt=0%2C5&q=LANUV+NRW+Investigations+on+the+input+and+elimination+of+hazardous+substances+in+sewage+treatment+plants&btnG=(accessed on 14 June 2022) Umweltbundesamt Maßnahmen zur Verminderung des Eintrages von Mikroschadstoffen in die Gewässer—Phase 2. Texte 60Available online: https://www.umweltbundesamt.de/sites/default/files/medien/377/publikationen/mikroschadstoffen_in_die_gewasser-phase_2.pdf2016(accessed on 14 June 2022) GeissenV. MolH. KlumppE. UmlaufG. NadalM. Van Der PloegM. van de ZeeS.E.A.T.M. RitsemaC.J. Emerging pollutants in the environment: A challenge for water resource managementInt. Soil Water Conserv. Res.20153576510.1016/j.iswcr.2015.03.002 WandaE. NyoniH. MambaB. MsagatiT. Occurrence of emerging micropollutants in water systems in Gauteng, Mpumalanga, and North West Provinces, South AfricaInt. J. Environ. Res. Public. Health2017147910.3390/ijerph1401007928098799PMC5295330 AhmedM.B. JohirM.A.H. NgoH.H. GuoW. ZhouJ.L. BelhajD. MoniM.A. Chapter 4—Methods for the analysis of micro-pollutantsCurrent Developments in Biotechnology and Bioengineering VarjaniS. PandeyA. TyagiR.D. NgoH.H. LarrocheC. ElsevierAmsterdam, The Netherlands202010.1016/B978-0-12-819594-9.00004-89780128195949 SchwarzenbachR.P. EscherB.I. FennerK. HofstetterT.B. JohnsonC.A. von GuntenU. WehrliB. The challenge of micropollutants in aquatic systemsScience20063131072107710.1126/science.112729116931750 KumarN.M. SudhaM.C. DamodharamT. VarjaniS. Micro-pollutants in surface water: Impacts on the aquatic environment and treatment technologiesCurrent Developments in Biotechnology and BioengineeringEmerging Organic Micro-pollutantsElsevierAmsterdam, The Netherlands2020Chapter 3416210.1016/B978-0-12-819594-9.00003-6 WichernM. HeinzE. Abwassertechnisches Laborpraktikum “Forschungspraktikum Siedlungswasserwirtschaft”. Lehrstuhl für Siedlungswasserwirtschaft und Umwelttechnik, Ruhr Universität Bochum. Abwassertechnisches Laborpraktikum Forschungspraktikum Siedlungswasserwirtschaft—PDF Free Download2017Available online: docplayer.org(accessed on 14 June 2022) AlvarinoT. SuarezS. LemaJ. OmilF. Understanding the sorption and biotransformation of organic micropollutants in innovative biological wastewater treatment technologiesSci. Total Environ.201761529730610.1016/j.scitotenv.2017.09.27828982079 ChavoshaniA. HashemiA. AminM.M. AmetaS.C. Micropollutants and Challenges: Emerging in the Aquatic Environments and Treatment ProcessesElsevierAmsterdam, The Netherlands2020Chapter 110.1016/B978-0-12-818612-1.00007-6978-0-12-818612-1 TijaniJ. FatobaO. PetrikL.F. A Review of pharmaceuticals and endocrine-disrupting compounds: Sources, effects, removal, and detectionsWater Air Soil Pollut.2013224117010.1007/s11270-013-1770-3 IKSR Auswertungsbericht Biozide und Korrosionsschutzmittel. Nr. 183Internationale Kommission zum Schutze des RheinsKoblenz, Germany20103-941994-15-8 AbegglenC. EscherB. HollenderJ. KoepkeS. OrtC. PeterA. SiegristH. von GuntenU. ZimmermannS.G. KochM. Ozonung von Gereinigtem Abwasser. Schlussbericht Pilotversuch Regensdorf. Studie der Eawag im Auftrag des BAFU und des AWEL. Eidgenössische Anstalt für Wasserversorgung, Abwasserreinigung und Gewässerschutz. Zürich2009Available online: http://spurenstoffelimination.de/files/Ozonung_Abwasser_Schlussbericht_Regensdorf.pdf(accessed on 4 April 2022) MieheU. Wirksamkeit technischer Barrieren zur Entfernung von anthropogenen Spurenstoffen—Kläranlagen und RaumfiltPh.D. DissertationTechnischen Universität BerlinBerlin, Germany2016 BöhlerM. SiegristH. Aktivkohledosierung in den Zulauf zur Flockungs-Sandfiltration Kläranlage Kloten/Opfikon—Grosstechnische Umsetzung im Rahmen ergänzender Untersuchungen zum MicroPoll-Projekt. DWA-Seminar Weitergehende Abwasserreinigung. Bonn, 18.01.2011Maßnahmen zur Verminderung des Eintrages von Mikroschadstoffen in die Gewässer—Phase 2Texte 60/2016UBADessau-Roßlau, Germany2016 GrünebaumT. Elimination von Arzneimittelrückständen in kommunalen Kläranlagen. Elimination von Arzneimitteln und organischen Spurenstoffen: Entwicklung von Konzeptionen und innovativen, kostengünstigen Reinigungsverfahren. Schlussbericht Phase 1. Düsseldorf: Ministerium für Klimaschutz, Umwelt, Landwirtschaft, Natur- und Verbraucherschutz des Landes Nordrhein-WestfalenMaßnahmen zur Verminderung des Eintrages von Mikroschadstoffen in die Gewässer—Phase 2Texte 60/2016UBADessau-Roßlau, Germany2016 BornemannC. ErbeV. HachenbergM. KolischG. OsthoffT. TaudienY. Einsatz von Pulveraktivkohle in vorhandene Flockungsfiltrationsanlagen am Beispiel der Kläranlage Buchenhofen. Tagungsband, Gesellschaft zur Förderung der Siedlungswasserwirtschaft an der RWTH Aachen (45. Essener Tagung für Wasser- und Abfallwirtschaft)Maßnahmen zur Verminderung des Eintrages von Mikroschadstoffen in die Gewässer—Phase 2Texte 60/2016UBADessau-Roßlau, Germany2016 CraneM. WattsC. BoucardT. Chronic aquatic environmental risks from exposure to human pharmaceuticalsSci. Total Environ.2006367234110.1016/j.scitotenv.2006.04.01016762401 MalajE. von der OheP.C. GroteM. KühneR. MondyC.P. Usseglio-PolateraP. BrackW. SchäferR.B. Organic chemicals jeopardize the health of freshwater ecosystems on the continental scaleProc. Natl. Acad. Sci. USA20141119549955410.1073/pnas.132108211124979762PMC4084479 PrasseC. StalterD. Schulte-OehlmannU. OehlmannJ. TernesT.A. Spoilt for choice: A critical review on the chemical and biological assessment of current wastewater treatment technologiesWater Res.20158723727010.1016/j.watres.2015.09.02326431616 DiehleM. GebhardtW. PinnekampJ. SchäfferA. LinnemannV. Ozonation of valsartan: Structural elucidation and environmental properties of transformation productsChemosphere201921643744810.1016/j.chemosphere.2018.10.12330384314 JiangJ. ZhouZ. SharmaV.K. Occurrence, transportation, monitoring and treatment of emerging micro-pollutants in waste water — A review from global viewsMicrochem. J.201311029230010.1016/j.microc.2013.04.014 O’FlynnD. LawlerJ. YusufA. Parle-McDermottA. HaroldD. Mc CloughlinT. HollandL. ReganF. WhiteB. A review of pharmaceutical occurrence and pathways in the aquatic environment in the context of a changing climate and the COVID-19 pandemicAnal. Methods20211357510.1039/D0AY02098B33507166 DaughtonC.G. TernesT.A. Pharmaceuticals and personal care products in the environment: Agents of subtle change?Environ. Health Perspect.1999107(Suppl. S6)90793810.1289/ehp.99107s690710592150PMC1566206 JjembaP.K. Excretion and ecotoxicity of pharmaceutical and personal care products in the environmentEcotoxicol. Environ. Saf.20066311313010.1016/j.ecoenv.2004.11.01116399163 PetrieB. BardenR. Kasprzyk-HordernB. A review on emerging contaminants in wastewaters and the environment: Current knowledge, understudied areas and recommendations for future monitoringWater Res.20157232510.1016/j.watres.2014.08.05325267363 SchmidtT.C. Recent trends in water analysis triggering future monitoring of organic micropollutantsAnal. Bioanal. Chem.20184103933394110.1007/s00216-018-1015-929564501PMC6010479 HelmenstineA.-M. Spectroscopy Introduction. Thought Co. 26 August 2020Available online: https://www.thoughtco.com/introduction-to-spectroscopy-603741(accessed on 22 April 2022) Sapkota, Anupama 22 Types of Spectroscopy with Definition, Principle, Steps, Uses12January2022Available online: https://microbenotes.com/types-of-spectroscopy/(accessed on 23 April 2022) KieratR.M. Synthese, Modifikation und Biologische Anwendung Fluoreszierender XanthenfarbstoffeInaugural DissertationRuprechts-Karls-Universität HeidelbergHeidelberg, Germany2008Available online: http://www.ub.uni-heidelberg.de/archiv/8301(accessed on 28 January 2022) SauerM. HofkensJ. EnderleinJ. Handbook of Fluorescence Spectroscopy and Imaging: From Single Molecules to EnsemblesWileyHoboken, NJ, USA201110.1002/9783527633500 LakowiczJ.R. Principles of Fluorescence SpectroscopySpringerLondon, UK201010.1007/978-0-387-46312-4 ChinenA. GuanC. FerrerJ. BarnabyS. MerkelT. MirkinC. Nanoparticle Probes for the Detection of Cancer Biomarkers, Cells, and Tissues by FluorescenceChem. Rev.2015115105301057410.1021/acs.chemrev.5b0032126313138PMC5457709 SchraderB. Chemische Anwendungen der Raman-SpektroskopieAngew. Chem.19738592595010.1002/ange.19730852102 VandenabeeleP. Practical Raman spectroscopy: An introductionThe Atrium Southern Gate Chichester West Sussex United KingdomWileyHoboken, NJ, USA2013 PostC. BrülisauerS. WaldschlägerK. HugW. GrüneisL. HeydenN. SchmorS. FördererA. ReidR. ReidM. Application of Laser-Induced, Deep UV Raman Spectroscopy and Artificial Intelligence in Real-Time Environmental Monitoring—Solutions and First ResultsSensors202121391110.3390/s2111391134198916PMC8201312 AsherS.A. JohnsonC.R. A new selective technique for characterization of polycyclic aromatic hydrocarbons in complex samples: UV resonance Raman spectroscopy of coal liquidsAnal. Chem.19845622582261650787210.1021/ac00276a065 LiuC. Implementation of Deep Ultraviolet Raman SpectroscopyPh.D. ThesisTechnical University of DenmarkKongens Lyngby, Denmark2011 Photon Systems Inc. DUV Raman PL200. Fully Integrated, Lab Model Deep UV Resonance Raman & Photoluminescence Spectrometer, with Microscopic Imaging2017Available online: https://photonsystems.com/wp-content/uploads/2019/02/DUV-RamanPL200-Data-Sheet-V12-Web.pdf(accessed on 10 February 2022) Photon Systems Inc. New Deep UV Raman & Photoluminescence Spectrometer System: The DUV Raman/PL 200Proceedings of the SPIE DCS 2018Orlando, FL, USA17–19 April 2018 Photon Systems Inc. Fused Fluorescence and Raman—Photon Systems, 9.3.20202017Available online: https://photonsystems.com/fused-fluorescence-and-Raman/(accessed on 10 February 2022) HendersonR.K. BakerA. MurphyK.R. HamblyA. StuetzR.M. KhanS.J. Fluorescence as a potential monitoring tool for recycled water systems: A reviewWater Res.20094386388110.1016/j.watres.2008.11.02719081598 SpencerR.G.M. BoltonL. BakerA. Freeze/thaw and pH effects on freshwater dissolved organic matter fluorescence and absorbance properties from a number of UK locationsWater Res.2007422941295010.1016/j.watres.2007.04.01217540432 BakerA. CumberlandS.A. BradleyC. BuckleyC. BridgemanJ. To what extent can portable fluorescence spectroscopy be used in the real-time assessment of microbial water quality?Sci. Total Environ.2015532141910.1016/j.scitotenv.2015.05.11426057622 ReynoldsD.M. AhmadS.R. The effect of metal ions on the fluorescence of sewage wastewaterWater Res.1995292214221610.1016/0043-1354(95)00046-NBMUV Kurzinfo Abwasser—KläranlageAvailable online: https://www.bmuv.de/en/topics/water-resources-waste/water-management/wastewater/sewage-treatment-plant(accessed on 4 April 2022) SonnenburgA. MosbachJ. GrimmelO. UrbanW. Leistungssteigerung von SandfängenWien. Mitt.202025291112Available online: https://repositum.tuwien.at/bitstream/20.500.12708/15607/1/sonnenburg_leistungssteigerung%20von%20sandf%c3%a4ngen.pdf(accessed on 3 February 2022) EngelhardtM. WagnerM. BehnischJ. BlachT. SchwarzM. AbwassertechnikHandbuch für Bauingenieure ZilchK. DiederichsC.J. BeckmannK.J. GertzC. MalkwitzA. MoormannC. Springer FachmedienWiesbaden, Germany2020 GuptaV.K. AliI. SalehT.A. NayakaA. AgarwalS. Chemical treatment technologies for waste-water recycling—An overviewRSC Adv.201226380638810.1039/c2ra20340e FranklinL.B. Wastewater Engineering: Treatment. Disposal and ReuseMcGraw Hill, Inc.New York, NY, USA1991 RosenwinkelK. KroissH. DichtlN. SeyfriedC. WeilandP. AnaerobtechnikSpringerBerlin/Heidelberg, Germany2015 MelinT. JeffersonB. BixioD. ThoeyeC. De WildeW. De KoningJ. van der GraafJ. WintgensT. Membrane bioreactor technology for wastewater treatment and reuseDesalination200618727128210.1016/j.desal.2005.04.086 FörstnerU. KösterS. UmweltschutztechnikSpringerBerlin/Heidelberg, Germany2018 WolfY. OsterS. ShuliakevichA. BrücknerI. DolnyR. LinnemannV. PinnekampJ. HollertH. SchiwyS. Improvement of wastewater and water quality via a full-scale ozonation plant?—A comprehensive analysis of the endocrine potential using effect-based methodsSci. Total Environ.202280314975610.1016/j.scitotenv.2021.14975634492496 RadjenovićJ. PetrovićM. BarcelóD. Advanced mass spectrometric methods applied to the study of fate and removal of pharmaceuticals in wastewater treatmentTrends Anal. Chem.2007261132114410.1016/j.trac.2007.10.002 FondermannP. Investigation of Correlation of Selected Amino Acids to Sum Parameters in Monitoring of Wastewater Treatment Plants Using Fluorescence Laser SpectrometryMaster’s ThesisDepartment of Engineering Geology and Hydrogeology, RWTH Aachen UniversityAachen, Germany11May2021 PinnekampJ. Grundlagen der Gewässergütewirtschaft und Siedlungswasserwirtschaft. Vorlesungsskript. RWTH Aachen, Lehrstuhl für Siedlungswasserwirtschaft und Wassergütewirtschaft (ISA)2019Available online: https://moodle.rwth-aachen.de/pluginfile.php/377447/mod_resource/content/0/GGSW_SS19_V02%20Charakterisierung%20Wasser%20Abwasser_190409(accessed on 11 April 2021)not publicly accessible PennM.R. PauerJ.J. MihelcicJ.R. Biochemical Oxygen Demand. Encyclopedia of Live Support Systems (EOLSS)2009Available online: http://www.eolss.net/sample-chapters/c06/e6-13-04-03.pdf(accessed on 9 April 2021)ISO 15705:2002Water Quality—Determination of the Chemical Oxygen Demand Index (ST-COD)—Small-Scale Sealed-Tube MethodISOGeneva, Switzerland2002ISO 6060:1989Water Quality—Determination of the Chemical Oxygen DemandISOGeneva, Switzerland1989ISO 3199Ausgabe 1975-02 Natriumchlorat für Industriezwecke; Bestimmung des Chloratgehaltes; Titrimetrische Dichromat-MethodeISOGeneva, Switzerland1975 ReinekeW. SchlömannM. Umweltmikrobiologie3rd ed.Springer SpektrumBerlin, Germany2020DIN EN 1484Water Analysis—Guidelines for the Determination of Total Organic Carbon (TOC) and Dissolved Organic Carbon (DOC)Beuth VerlagBerlin, Germany2019DIN EN 15216:2021-12Environmental Solid Matrices—Determination of Total Dissolved Solids (TDS) in Water and EluatesBeuth VerlagBerlin, Germany2021 KöppelS. DennerM. DOC/TOC im Eluat von Abfällen aus der MBA. Evaluierung der Bestimmungsverfahren. Wien: Umweltbundesamt GmbH (Report/Umweltbundesamt, [N.F.], p 298)2010Available online: https://www.umweltbundesamt.at/fileadmin/site/publikationen/REP0298.pdf(accessed on 29 November 2021) PineC.S.F. Process gas chromatographyTalanta19671426929710.1016/0039-9140(67)80001-818960097 VillalobosR. Process gas chromatographyAnal. Chem.197547983A1004A10.1021/ac60361a778 ClemonsJ. Chromatography in process analysisEncyclopedia of Analytical ChemistryWileyHoboken, NJ, USA201612710.1002/9780470027318.a2107 KesslerR.W. ProzessanalytikWileyWeinheim, Germany2006 StraubA. Einfache Messmethoden zur Charakterisierung sowie Maßnahmen zur Erhöhung der Zuverlässigkeit und Leistungsfähigkeit biologischer KleinkläranlagenPh.D. DissertationLehrstuhl Wassertechnik und Siedlungswasserbau der Brandenburgischen Technischen Universität CottbusCottbus, Germany2008 EPA Method 180.1:1993—Determination of Turbidity by NephelometryEPAWashington, DC, USA1993ISO 7027-1:2016(en)Water Quality—Determination of Turbidity—Part 1: Quantitative MethodsISOGeneva, Switzerland2016Available online: https://www.iso.org/standard/62801.html(accessed on 14 June 2022) ShiZ. ChowC.W.K. FabrisR. LiuJ. JinB. Applications of Online UV-Vis Spectrophotometer for Drinking Water Quality Monitoring and Process Control: A ReviewSensors202222298710.3390/s2208298735458971PMC9024714 WortbergM. KurzJ. Analytics 4.0: Online wastewater monitoring by GC and HPLCAnal. Bioanal. Chem.20194116783679010.1007/s00216-019-02065-w31402422 Photon Systems Inc. Manual: Raman PL 200Photon SystemsCovina, CA, USA2020 unpublished ParthierR. Messtechnik—Vom SI-Einheitensystem über Bewertung von Messergebnissen zu Anwendungen der Elektrischen Messtechnik9th ed.SpringerBerlin/Heidelberg, Germany2020978-3-658-27130-5KnowItAll-Software, Spektroskopie-EditionAvailable online: https://sciencesolutions.wiley.com/knowitall-software-spektroskopie-edition/?lang=de(accessed on 5 April 2022) GörgensM. Method Comparison of Laser Induced Fluorescence and Raman Spectroscopy Using Water Management Relevant MicropollutantsMaster’s ThesisDepartment of Engineering Geology and Hydrogeology, RWTH Aachen UniversityAachen, Germany6May2021 WongburiP. ParkJ.K. Big Data Analytics from a Wastewater Treatment PlantSustainability2021131238310.3390/su132212383 LiuJ. OsadchyM. AshtonL. FosterM. SolomoneC. GibsonS. Deep convolutional networks for Raman spectrum recognition: A unified solutionAnalyst20171424067407410.1039/C7AN01371J28993828Directive (EU) 2020/2184 of the European Parliament and of the Council of 16 December 2020 on the Quality of Water Intended for Human ConsumptionAvailable online: https://eur-lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32020L2184&from=ES(accessed on 22 April 2022) BhartiaR. HugW. ReidR. Improved sensing using simultaneous deep UV Raman and fluorescence detectionChemical, Biological, Radiological, Nuclear, and Explosives (CBRNE) Sensing XIII (83581A)International Society for Optics PhotonicsBellingham, WA, USA2012Volume 8358 ReinartzA. Investigation of Selected Disinfection Byproducts (DNP) in Ozonation of Wastewater Treatment Plants by Laser-Induced Raman and Fluorescence Spectroscopy—Possibilities and Limitations of ApplicationMaster’s ThesisDepartment of Engineering Geology and Hydrogeology, RWTH Aachen UniversityAachen, Germany31March2022 AhmadS.R. ReynoldsD.M. Synchronous fluorescence spectroscopy of wastewater and some potential constituentsWater Res.1995201599160210.1016/0043-1354(94)00266-A BridgemanJ. BakerA. Carliell-MarquetC. CarsteaE.M. Determination of changes in wastewater quality through a treatment works using fluorescence spectroscopyEnviron. Technol.2013343069307710.1080/09593330.2013.80313124617065 CarsteaE.M. PopaC.L. BakerA. BridgemanJ. In situ fluorescence measurements of dissolved organic matter: A reviewSci. Total Environ.202069913436110.1016/j.scitotenv.2019.13436131683216 SharpeT.J. Assessing a Fluorescence Spectroscopy Method for In-Situ Microbial Drinking Water QualityPh.D. DissertationPortland State UniversityPortland, OR, USA201710.15760/etd.5732 CumberlandS. BridgemanJ. BakerA. SterlingM. WardD. Fluorescence spectroscopy as a tool for determining microbial quality in potable water applicationsEnviron. Technol.20123368769310.1080/09593330.2011.58840122629644 MohrdieckM. Optimierung von Wasseraufbereitungsprozessen über ein Echtzeit-Monitoring von Summenparametern mit Hilfe der Fluoreszenz-LaserspektroskopieMaster’s ThesisDepartment of Engineering Geology and Hydrogeology, RWTH Aachen UniversityAachen, Germany31March2021 BaldwinD.S. ValoW. Exploring the relationship between the optical properties of water and the quality and quantity of dissolved organic carbon in aquatic ecosystems: Strong correlations do not always mean strong predictive powerEnviron. Sci. Process. Impacts20151761963010.1039/C4EM00473F25623568 KowalczukP. CooperW.J. DurakoM.J. KahnA.E. GonsiorM. YoungH. Characterization of dissolved organic matter fluorescence in the South Atlantic Bight with use of PARAFAC model: Relationships between fluorescence and its components, absorption coefficients and organic carbon concentrationsMar. Chem.2010118223610.1016/j.marchem.2009.10.002 ShutovaY. BakerA. BridgemanJ. HendersonR.K. Spectroscopic characterization of dissolved organic matter changes in drinking water treatment: From PARAFAC analysis to online monitoring wavelengthsWater Res.20145415910.1016/j.watres.2014.01.05324568785 SchmorS. Development of Raman Measurement Standards for Selected Substances in Sewage—Conception and Execution of a Flow ExperimentMaster’s ThesisDepartment of Engineering Geology and Hydrogeology, RWTH Aachen UniversityAachen, Germany30April2020 RadhakrishnaS. KarguppikarA.M. Raman Spectra of Irradiated Alkali Bromate CrystalsJ. Phys. Soc. Jpn.1973257858110.1143/JPSJ.35.578

Material flow analysis of micropollutants in the environment (adapted with permission from [19]. Copyright 2010, IKSR—CIPR—ICBR—all rights reserved).

Wavelength shift due to Stokes shift, (a); Jablonski diagram, (b) (adapted with permission from [40]. Copyright 2015, American Chemical Society).

Optical setup of the Deep-UV Raman and Photoluminescence 200 Spectrometer visualized with Wondershare EdrawMax Copyright © 2022 Edrawsoft.

Raman calibration using Acetonitrile (Spectrum Analyzer) measured with 30 pulses (a). Fluorescence calibration using a GaN-sample (Spectrum Analyzer), measured with 1 pulse, (b).

Fluorescence spectra of four carbamazepine concentrations, averaged from 10 repeated measurements. The spectra were scaled and the ethanol spectrum was subtracted. The spectrum of the solvent ethanol (dotted line black), measured in a quartz cuvette and the empty cuvette itself (dotted line grey) are shown. Settings: pulse number 50; pulse frequency 40; slit size 150 µm; grating 300 ln/mm; focal length 20 mm (raw data from [85]).

Influence of the measurement set-up on the signal: Raman spectrum of two nitrate solutions with 50 mg/L (a) and 1 mg/L NO3- (b) in a cuvette (red line) and in a flow cell (black line). Critical interpretation arises at the nitrate reference peak at around 1050 1/cm because cuvette and nitrate signal appear at the same wavenumber, indicating the presence of higher nitrate concentrations. Settings: pulse number 500; pulse frequency 40/s; slit size 150 µm; grating 3600 ln/mm; focal length 20 mm (only cuvette) (raw data from [90]).

Overview of the extraction points of the 24-h collective samples of two different wastewater treatment plants and their targets around the biological treatment (aeration tank and membrane bioreactor) and the ozonation.

Averaged and scaled fluorescence analysis of 7 wastewater samples after the treatment in a membrane bioreactor [96] and as reference, DOC concentrations of the same samples in [mg/L] analyzed by a collaborating laboratory. Settings: pulse number 20; pulse frequency 40/s; slit size 150 µm; grating 300 ln/mm; focal length 20 mm, measured in a quartz glass cuvette (raw data from [96]).

Correlation curve, functional equation, and coefficient of determination between DOC and fluorescence peak intensity from 7 effluent samples exiting the membranes bioreactor from two different WWTPs (raw data from [96]).

Dilution series from averaged and scaled tryptophan spectra (a) after subtraction of solvent spectrum. Due to oversaturation of the CCD detector the 10 mg/L and 5 mg/L concentrated solution spectra are cut off. In the Y-axis enlargement (b) the detection limit of tryptophan is clearly seen at the signal of 0.001 mg/L (green line). Settings: pulse number 10, pulse frequency 40/s, focal length 20 mm, measured in a quartz glass cuvette (raw data from [63]).

Dilution series from averaged and scaled tyrosine spectra (a) after subtraction of solvent spectrum. Enlargement of the Y-axis emphasizes the detection limit of the substance (b). Settings: pulse number 10, pulse frequency 40/s, focal length 20 mm, measured in a quartz glass cuvette (raw data from [63]).

Correlation curve, functional equation, and coefficient of determination between DOC and the intensities of fluorescence spectra of peak T2 of all influent and effluent samples around the aeration tank (raw data from [63]).

Comparison between the fluorescence spectra of WWTP influent (inf.) (solid lines) and effluent (eff.) (dotted lines) samples of the same sampling day; measurement results after complete processing including subtract spectrum. The reddish background marks the characteristic tyrosine and tryptophan peak areas; the grey background represents the characteristic peaks of DOC and BOD. Influent and effluent samples with sample number and content of DOC (in square brackets), determined in a collaborating laboratory. Settings: pulse number 10; slit size 150 µm; grating 300 ln/mm; focal length 20 mm; pulse frequency 40/s (raw data from [63]).

A total of 10 repeated measurements of fluorescence of metformin hydrochloride as a powder, showing slight photodegradation phenomenon. Settings: pulses number 200, grating 300 ln/mm, slit seize 150 µm, pulse frequency 40/s, measured on a turning table (raw data from [85]).

Averaged Raman (a) and fluorescence (b) spectra of carbamazepine before subtracting ethanol spectrum and scaling, including the specific spectrum of the empty cuvette and the solvent ethanol in the same cuvette. Settings above: pulse number 500, slit seize 150 µm, grating 3600 ln/mm, focal length 20 mm (in cuvette); settings bottom: pulse number 50, slit seize 150 µm, grating 300, focal length 20 mm (in cuvette) (raw data from [85]).

Averaged, scaled fluorescence spectra of naproxen after subtraction of ethanol. The measurements also emphasize the influence of the number of pulses for cuvette and ethanol. Settings: pulse number 1/10, slit seize 15 µm, grating 300, focal length 20 mm (raw data from [85]).

Raman measurements of dilution series of averaged nitrate spectra, measured in a flow cell to avoid cuvette disturbance (a) and a detailed enlargement of the nitrate reference peak at 1050 1/cm (b); settings: pulse number 500/1000, slit seize 150 µm, grating 3600 ln/mm, pulse frequency 40/s (raw data from [90]).

Correlation curve, functional equation and coefficient of determination between nitrate concentrations and Raman peak intensities referring to the dilution series of nitrate (raw data from [90]).

Raman measurements of dilutions series of averaged nitrite spectra, measured in a flow cell to avoid cuvette disturbance (a) and a detailed enlargement of the nitrite reference peak at 1325 1/cm (b); settings: pulse number 500/1000, slit seize 150 µm, grating 3600 ln/mm, pulse frequency 40/s (raw data from [90]).

Averaged Raman bromate spectra of the dilution series with 6 different concentrations (a) and detailed enlargement of the enveloping curve of the water signal from 400 to 800 1/cm and its low effect of different bromate concentrations (b). Settings: pulse number 500/1000, slit seize 150 µm, grating 3600, pulse frequency 40/s (raw data from [90]).

Raman reference spectrum of solid potassium bromate (KBrO3) from the spectrabase.com database and table with Raman active signals of solid bromate compounds and dissolved bromate anions, according to [101]. (Adapted with permission from [101]. Copyright 1973, Journal of the Physical Society of Japan (JPSJ)).

sensors-22-04668-t001_Table 1

Measured substances with manufacturer information, concentrations of the dilution series and given limits.

Substance; CAS no.ManufacturerMeasurementConcentrationLegal Limits
Metformin hydrochloride; 1115-70-4; purity > 99%BioTrend, Cologne, GermanyPure powderin solution with cuvette from Starna Cells®Atascadero, CA, USAPure substance and 10; 1; 0.1; 0.01 g/L in solution with Millipore®, Merck KGaADarmstadt, GermanyPreventive value 0.1 µg/L °
Carbamazepine; 297-46-4; purity > 98%BioTrend, Cologne, GermanyPure powderin solution with cuvette from Starna Cells®Atascadero, CA, USAPure substance and 10; 1; 0.1; 0.01 mg/g in solution with ethanolPreventive value 0.1 µg/L °
Hydrochlorothiazide; 58-93-5; purity > 99%BioTrend, Cologne, GermanyPure powder in solution with cuvette from Starna Cells®Atascadero, CA, USAPure substance and 10; 1; 0.1; 0.01 mg/g in solution with ethanolPreventive value 0.1 µg/L °
Acetaminophen; 103-90-2; purity > 99%BioTrend, Cologne, GermanyPure powderin solution with cuvette from Starna Cells®Atascadero, CA, USAPure substance and 10; 1; 0.1; 0.01 mg/g in solution with ethanolPreventive value 0.1 µg/L °
Naproxen; 22204-53-1;purity > 98%BioTrend,Cologne,GermanyPure powderin solution with cuvette from Starna Cells®Atascadero, CA, USAPure substance and 0.005; 0.001; 0.0005; 0.00001 down to 1 × 10−5 mg/g in solution with ethanolPreventive value 0.1 µg/L °
Diclofenac; 15307-93-4;purity > 98%BioTrend,Cologne,GermanyPure powderin solution with cuvette from Starna Cells®Atascadero, CA, USAPure substance and 10; 5; 1; 0.5; 0.1; 0.05; 0.01 g/L in solution with Millipore®, Merck KGaA, Darmstadt, GermanyWater hazard class 3 + acc. to WDF watch list X
Bromate; 7789-38-0VWR ChemicalsIn solution with solvent water, in flow cell50; 25; 10; 5; 1; 0.1; 0.01 mg/L in solution with Millipore®, Merck KGaA, Darmstadt, Germany10 µg/L, according to Directive EU #
Bromide; 7647-15-6Supelco®,Merck KGaA, Darmstadt, GermanyIn solution with solvent water, in flow cell50; 25; 10; 5; 1; 0.1; 0.01 mg/L in solution with Millipore®Merck KGaA, Darmstadt, Germany-
Nitrate; 7631-99-4ROTI®StarCarl Roth GmbH + Co. KG,Karlsruhe, GermanyIn solution with solvent water, in flow cell50; 25; 10; 5; 1 mg/L in solution with Millipore®,Merck KGaA, Darmstadt, Germany50 mg/L, according to WFD X and Directive EU #
Nitrite; 7632-00-0Certipur®Merck KGaA,Darmstadt, GermanyIn solution with solvent water, in flow cell50; 25; 10; 5; 1 mg/L in solution with Millipore®,Merck KGaA, Darmstadt, Germany0.5 mg/L, according to Directive EU #
Tryptophan; 73-22-3Merck KGaA,Darmstadt, GermanyIn solution and as a solid, cuvette from Starna Cells®Pure substance and 10; 5; 1; 0.5; 0.1; 0.05 0.01; 0.005; 0.001 mg/L in solution with Millipore®,Merck KGaA, Darmstadt, Germany-
Tyrosine; 60-18-4Carl Roth®Carl Roth GmbH + Co. KG,Karlsruhe, GermanyIn solution and as a solid, cuvette from Starna Cells®Pure substance and 10; 5; 1; 0.5; 0.1; 0.05 0.01; 0.005; 0.001 mg/L in solution with Millipore®,Merck KGaA, Darmstadt, Germany-

° Preventive value, according to LANUV 2015 and https://www.flussgebiete.nrw.de/monitoringleitfaden-oberflaechengewaesser-anhang-d4-7724 (accessed on 7 February 2022); + Water Resources Act and https://echa.europa.eu/documents/10162/13641/rest_microplastics_axvreport_annex_en.pdf/01741d07-f06b-bf32-8d6f-d6a8de54c4d0 (accessed on 7 February 2022). Administrative regulation on substances hazardous to water; https://echa.europa.eu/documents/10162/13641/rest_microplastics_axvreport_annex_en.pdf/01741d07-f06b-bf32-8d6f-d6a8de54c4d0 (accessed on 7 February 2022). # Directive (EU) 2020/2184 of the European Parliament and of the Council of 16 December 2020 on the quality of water intended for human consumption, https://eur-lex.europa.eu/legal-content/EN/TXT/PDF/?uri=CELEX:32020L2184&from=ES (accessed on 22 April 2022) [88]; X WFD Directive 2000/60/EC of the European Parliament and of the Council of 23 October 2000 establishing a framework for Community action in the field of water policy, https://eur-lex.europa.eu/resource.html?uri=cellar:5c835afb-2ec6-4577-bdf8-756d3d694eeb.0004.02/DOC_1&format=PDF (accessed on 22 April 2022).

sensors-22-04668-t002_Table 2

Spectral analysis of the diluted substances and their detection rate (* low; ** medium; *** high; - undetectable), measured by the spectrometer used.

Substance; CAS no.Spectral AnalysisDetection
RamanFluorescenceRamanFluorescence
Metformin hydrochloride; 1115-70-4; purity > 99% ✓✓--
Carbamazepine; 297-46-4; purity > 98% ✓✓***
Hydrochlorothiazide; 58-93-5; purity > 99%✓✓***
Acetaminophen; 103-90-2; purity > 99% ✓✓***
Naproxen; 22204-53-1; purity > 98% ✓✓*****
Diclofenac; 15307-93-4; purity > 98%✓✓--
Bromate; 7789-38-0 ✓✓-**
Bromide; 7647-15-6✓✓-**
Nitrate; 7631-99-4✓✓*-
Nitrite; 7632-00-0✓✓*-
Tryptophan; 73-22-3✓✓******
Tyrosine; 60-18-4✓✓*****
DOC ✕✓-**

High detection rate down to 0.001 mg/L; medium detection rate down to 0.01 mg/L; low detection rate down to 0.1 mg/L. (✓) measured; (✕) not measured.